Biochemical incubator related introduction 1. Purpose The intelligent biochemical incubator in the experiment to do bacteria, mold, micro-organisms, special thermostat equipment for preservation. Through the standardized operation of the biochemical incubator to obtain accurate results, it is in a good working condition and ensures the health and safety of the operators.
2. The scope applies to the operation, maintenance and maintenance of intelligent biochemical incubators.
3. Responsibility The laboratory technician conducts operations, repairs, maintenance and records.
The surveyor is responsible for its verification. The Health Inspection Division is responsible for supervision.
4. Operating Procedures 4.1 Normal Operating Conditions Relative Humidity: ≤ 80% RH Ambient Temperature: 5~32°C Pressure: 86~106Kpa There is no strong vibration and corrosive gas around the incubator. The incubator should avoid direct sunlight or other cold and heat sources.
4.2 How to use 4.2.1 Turn on the power supply consistent with the requirements of this instrument, and ground the power supply outlet reliably.
4.2.2 Turn on the power switch, the power indicator light is on, and the temperature controller starts to display the temperature in the working room.
4.2.3 Set the temperature according to the required temperature of the culture. After the set temperature is automatically constant, open the door and put it into the culture.
5. Precautions and routine maintenance 5.1 Firstly, the grounding work of the equipment should be done. Grounding should be done twice as thick as the power wire.
5.2 The bottom of the studio should not be placed on the test sample, the sample on the shelf should not be placed too dense, so as not to affect the flow of air circulation within the box.
5.3 When it is used for the first time or for a long time, it should be started for 6 to 8 hours without load to remove the malfunctions that may occur during transportation or storage before normal use to avoid unnecessary losses.
5.4 The biochemical tank must be placed in a place away from light, ventilation and drying. The distance between the equipment and the wall should not be less than 10cm.
5.5 It is forbidden to scrub the inner and outer surfaces with acid, alkali and other corrosive materials. The neutral detergent can be used for regular cleaning and the dry cloth can be wiped clean.
5.6 When the fuse is replaced, replace the same fuse according to the fuse current and the size of the fuse.
5.7 When the biochemical tank is not in use, the power should be cut off and the inside of the tank should be kept clean and dry.